Transcatheter aortic valve implantation in a patient with circulatory collapse, using the lucas® chest compression system

Peter Blom Jensen, Claus Andersen, Henrik Nissen

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

We describe a case of Transcatheter Aortic Valve Implantation (TAVI) using the LUCAS® Chest Compression System in an elderly high risk patient with severe aortic stenosis and heart failure. In this case, the patient developed severe aortic regurgitation following predilatation of the native aortic valve and automated cardiopulmonary resuscitation (A-CPR) was initiated. The procedure was performed under ongoing A-CPR for a total of 28 min. The patient was transferred to the intensive care unit and to a step down unit the following day. At follow-up 30 days later, she showed no signs of neurologic or cardiac damage. This case report shows, that it is possible to perform the TAVI procedure under ongoing A-CPR and that A-CPR, judged by invasive blood pressures, was capable of maintaining a satisfactory perfusion pressure even with a damaged aortic valve.

Original languageEnglish
JournalCatheterization and Cardiovascular Interventions
Volume81
Issue number6
Pages (from-to)1084-1086
ISSN1522-726X
DOIs
Publication statusPublished - May 2013

Keywords

  • cardiopulmonary support
  • complications adult cath/intervention
  • hemodynamics
  • valvular heart disease
  • ventricular support devices
  • Humans
  • Risk Factors
  • Aortic Valve Stenosis/diagnosis
  • Treatment Outcome
  • Balloon Valvuloplasty/adverse effects
  • Equipment Design
  • Shock/diagnosis
  • Cardiac Catheterization
  • Heart Valve Prosthesis Implantation/methods
  • Aged, 80 and over
  • Female
  • Hemodynamics
  • Cardiopulmonary Resuscitation/instrumentation

Fingerprint