TY - JOUR
T1 - Mechanical and Geometric Characterization of a Novel 2-Ply Vacuum-Pressed Biological Scaffold Patch Design for Posterior Mitral Valve Reconstruction
AU - Jedrzejczyk, Johannes H.
AU - Andersen, Frederik T.
AU - Petersen, Jacob
AU - Kaspersen, Alexander Emil
AU - Sahana, Urjosee
AU - Skov, Søren N.
AU - Væsel, Jens T.
AU - Hasenkam, J. Michael
AU - Tjørnild, Marcell J.
PY - 2024
Y1 - 2024
N2 - To assess the mechanical properties of small intestinal submucosal extracellular matrix (SIS-ECM) iterations and choose the optimal version for evaluating functional geometrics after posterior mitral valve reconstruction. Four SIS-ECM versions (2- and 4-ply vacuum-pressed and lyophilized) underwent uniaxial tensile testing. A posterior mitral valve reconstruction patch was developed based on MRI scans (n = 5). Posterior mitral valve reconstruction using 2-ply vacuum-pressed SIS-ECM was performed (n = 7), and geometrics were evaluated using a modified left heart simulator. The vacuum-pressed iterations displayed superior maximum stress values compared to lyophilized (2-ply: median [IQR], 15.8 [15.2–19.0] vs 7.9 [7.3–8.3] MPa, p < 0.001; 4-ply: median (IQR), 15.8 –[14.6–22.0] vs 7.9 [7.6–8.4] MPa). All reconstructed valves were competent with preserved total leaflet area, but individual leaflet segment areas were redistributed. Posterior mitral valve reconstruction with our 2-ply vacuum-pressed SIS-ECM patch design was feasible in vitro. Further in vivo evaluation is warranted. Graphical Abstract: (Figure presented.)
AB - To assess the mechanical properties of small intestinal submucosal extracellular matrix (SIS-ECM) iterations and choose the optimal version for evaluating functional geometrics after posterior mitral valve reconstruction. Four SIS-ECM versions (2- and 4-ply vacuum-pressed and lyophilized) underwent uniaxial tensile testing. A posterior mitral valve reconstruction patch was developed based on MRI scans (n = 5). Posterior mitral valve reconstruction using 2-ply vacuum-pressed SIS-ECM was performed (n = 7), and geometrics were evaluated using a modified left heart simulator. The vacuum-pressed iterations displayed superior maximum stress values compared to lyophilized (2-ply: median [IQR], 15.8 [15.2–19.0] vs 7.9 [7.3–8.3] MPa, p < 0.001; 4-ply: median (IQR), 15.8 –[14.6–22.0] vs 7.9 [7.6–8.4] MPa). All reconstructed valves were competent with preserved total leaflet area, but individual leaflet segment areas were redistributed. Posterior mitral valve reconstruction with our 2-ply vacuum-pressed SIS-ECM patch design was feasible in vitro. Further in vivo evaluation is warranted. Graphical Abstract: (Figure presented.)
KW - Lyophilized
KW - Mitral valve reconstruction
KW - Mitral valve repair
KW - Posterior mitral valve leaflet
KW - Small intestinal submucosal extracellular matrix
KW - Vacuum-pressed
U2 - 10.1007/s12265-024-10572-0
DO - 10.1007/s12265-024-10572-0
M3 - Journal article
C2 - 39466570
AN - SCOPUS:85207326720
SN - 1937-5387
JO - Journal of Cardiovascular Translational Research
JF - Journal of Cardiovascular Translational Research
ER -